Freeze-dryer EPSILON 2-6 LSCplus
7 Malfunctions and error correction
Version 10/2012, Rev. 1.6 of 16/12/2014 • sb
91
Translation of the original operating manual
7.1.2.2 Aeration valve, media drain valve
A malfunction of the aeration valve or the media drain valve may have
several causes. One potential source are contaminants such as product
residues within the valve.
Switch the freeze-dryer off and disconnect the mains plug.
Clean the valve (see chapter 8.1.1.3 - "Aeration valve, media drain
valve").
Put the freeze-dryer into operation again.
If there is still a leakage, the freeze-dryer must be checked by qualified
specialist personnel (see chapter 7.3 - "Service contact").
Pos: 181 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 182 /20 0 Chri st/3 60 GT-BA Labo r-Pi lot (STAND ARD MODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 20-0025 D ruc kste uerv entil @ 25\ mod_1 4049 840 63525_ 68.d ocx @ 184 209 @ 4 @ 1
7.1.2.3 Pressure control valve
A malfunction of the pressure control valve may have several causes.
NOTE
The inspection of the valve must be carried out by qualified specialist
personnel (see chapter 7.3 - "Service contact").
Pos: 183 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 184 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (S TANDA RDMOD U LE)/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 20- 0030 G ummiv entile @ 2 5\m od_1404 984 064 588 _68.doc x @ 18 422 3 @ 4 @ 1
7.1.2.4 Rubber valves
In order to identify a leaking rubber valve, the valves must be checked
individually:
Remove the rubber valve and seal the connection at the drying
chamber with a rubber stopper.
Check the tightness under vacuum until the leaking valve has been
localised.
Clean the valve or replace it if necessary.
Pos: 185 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 186 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (S TANDA RDMOD U LE)/0 70 Stö run gen und Fe hlers uche /07 0-0010 -00 20- 0040 Vaku um messs ond e @ 2 5\mod_ 140 498 4065 691_68 .docx @ 18 423 7 @ 4 @ 1
7.1.2.5 Vacuum sensor
Vacuum sensors have a limited service life and can be ordered as spare
parts (e.g. VSP62/63, order no. 125468).
Functional test:
Connect the vacuum sensor directly to the suction side connector of the
vacuum pump.
If a final pressure of at least 0.011 mbar is reached (with a vacuum pump
that has reached its operating temperature), the sensor and the vacuum
pump are OK.