Freeze-dryer EPSILON 2-6 LSCplus
7 Malfunctions and error correction
92
Version 10/2012, Rev. 1.6 of 16/12/2014 • sb
Translation of the original operating manual
Pos: 188 /20 0 Ch rist/3 61 G T-BA Labo r-P ilot (PR OJEK TE)/E psilo n S tanda rd/0 70 S törun gen und Fehl ers uch e/07 0-0010 -00 31 Eis ko ndens at or- und Stel lfläc hent emp era tur E psi lon @ 25\mo d_1 4049 846 659 83_ 68.doc x @ 185861 @ 3 @ 1
7.1.3 Insufficient ice condenser temperature
CAUTION
Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will be
impaired.
The refrigeration unit is equipped with a protective device against
overpressure in the refrigeration system and with a thermal motor
protection switch.
The protective devices trip
when the ambient temperature is too high
when the air circulation of the heat exchanger of the refrigeration
system is insufficient
when the refrigeration system is overloaded.
In these cases, the refrigeration unit will be switched off automatically. If the
permissible operating conditions are re-established after a cool-down
phase of several minutes, the refrigeration unit will be switched on again
automatically.
The malfunctions are displayed in the process and equipment information
window.
The minimum ice condenser temperature of approx. –85°C is reached
when the ice condenser is not loaded and the ice condenser chamber is
evacuated.
Pos: 189 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 190 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STA NDA RDMODU LE )/0 70 Stö run gen und Fehl ersuc he/ 070-0 021 P roz ess un d Anla gen meld ungen LS Cpl us @ 25\ mod_140 498 406 7666_6 8.doc x @ 1 842 65 @ 2 @ 1
7.2 Process and error messages
The LSCplus control system displays the complete process and error
messages, which is why they are not included in this operating manual.
You can order these documents from our service department.