Freeze-dryer EPSILON 2-6 LSCplus
8 Maintenance and service
100
Version 10/2012, Rev. 1.6 of 16/12/2014 • sb
Translation of the original operating manual
Pos: 218 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PR OJEK TE)/E psilo n Sta nda rd/0 80 Wa rtu ng und I nstan dhal tung /08 0-0 021 Wart ung splan Epsi lon 6er 10 er @ 25\ mod _14 049 8468138 5_6 8.d ocx @ 1859 31 @ 2 @ 1
8.3 Maintenance schedule
System/
Component
Detail / Part Repair/Rebuild
before
start
daily inspection replace operation calibration
Drying chamber:
dry and clean
X X
Steel bellow:
clean and undamaged
X X smooth travel
Silcone oil hoses and
X X leak free
Intermediate valve gasket
X X
after 50
sterilisations
Shelves: pins
X burr-free smooth travel
Shelves: surface
clean and undamaged
X
Shelves: guardrails
X
Acrylic door surface
X
fogging,
cracks
Door gasket:
clean and undamaged
X replace
after 50
sterilisations
Changing procedure
Door locks
X burr-free X Changing procedure
Hinges
X X X
Adjust hinges after re-
placing door gasket
Door cover:
securely attached
X X
Chamber sight glasses: clean,
illumination ok
X X
Electrical cabinet
X X
Leakages
X X
Noise emission during operation
X X
Dynamic gaskets: diaphragms
after 100
sterilisations
Static seals:
O-rings, clamp, KF
after 100
sterilisations
Loose wire
X
Cable tray closed
X
Framework
corrosion Remove corrosion
Aeration filter
after 50
sterilisations
Pump: oil level
X X X
Pump: oil condition
(colour, contamination)
X X X oil change
Oil mist filter:
emission of oil mist
X X X
Oil mist filter: oil level
X cleaning
Final vacuum of instrument
X
Residue / particles inside oil
casing
X
Oil leakages at pump
X X
after
10 years
Rotary shaft seal,
bearings,vanes, springs:
service life of 10 years
Vacuum leakage test
X
Inspection Periodical Maintenance
Drying
Chamber -
Operator Side
Machine Room
- General
Vacuum
System