Freeze-dryer EPSILON 2-6 LSCplus
5 Set-up and connection
38
Version 10/2012, Rev. 1.6 of 16/12/2014 • sb
Translation of the original operating manual
Fig. 20: Example of a door locking system of the loading door; here: door stop (left side)
Pos: 108 /01 0 Uni v ersal mod ule/L eerz ei le @ 0\m od_ 120 211624 450 0_0. doc x @ 11 4 @ @ 1
Pos: 109 /20 0 Ch rist/3 61 G T-BA L abo r-Pi lot (PRO JEK TE)/Eps ilo n Sta nda rd/0 50 Aufs tel lun g u nd Ansc hluss 0 50-006 0 Bel üft ungs ventil E psi lon @ 25\mo d_1 4049 846 327 70_68. doc x @ 185 777 @ 2 @ 1
5.6 Aeration valve
The aeration valve is located on top of the left side of the unit (see chapter
2.1.1 - "Functional and operating elements").
After the end of a freeze-drying process, the unit will be aerated via the
aeration valve.
NOTE
The ice condenser chamber can be flooded with nitrogen via the hose
nozzle of the aeration valve.
WARNING
For the aeration with gases under pressure, the client must install a safety
device for pressure relief.