Freeze-dryer EPSILON 2-6 LSCplus
6 Operation
Version 10/2012, Rev. 1.6 of 16/12/2014 • sb
59
Translation of the original operating manual
Click on touch: If this function is active, a clicking sound can be heard
whenever the system registers that a button has been touched.
Behaviour of the pressure control valve in the case of insufficient ice
condenser cooling: If this function is active (button "closed"), the pressure
control valve will close at an ice condenser temperature of
-20°C during the drying process in order to avoid damage to the vacuum
pump caused by the withdrawal of condensable gases. A corresponding
error message will be displayed. Pressing the button again deactivates the
function (button "controlled").
Access protection (4)
In this dialog box, the access rights can be managed on several levels and
they can be protected with a password.
In the factory setting with an activated access protection, data can be
viewed but not edited.
Fig. 47: Dialog box "Access protection"
NOTE
The other buttons cannot be activated unless the access protection is
active.
Password timer runtime: In order to prevent unauthorised access, the
system will automatically switch back to the default setting after a pre-
defined period of time.
In this case, there is a small lock symbol in the status line and below this
symbol the remaining time until the lock will be active is counted down. At
the same time a button with a big lock symbol will be displayed in the actual
values field.
Fig. 48: Countdown of the password timer and the button with the lock symbol